Abstract:
Two machines from two scientific centers (Russia and South Africa) were used for the manufacturing of
the Ti6Al4V alloys by the direct metal laser sintering. The chemical composition of powders complies
with the ASTM F-136 (grade 5), ASTM B348 (grade 23) standard for medical applications. Analysis of
the oxygen and nitrogen contamination in DMLS alloys was done with Van de Graaff accelerator with
two Mega Volts. It is found that structures of the samples manufactured with two different machines
used the same regimes are close to each other. TEM studies found the metastable martensitic structure
and silicon nitride Si3N4. It was found that the oxygen and nitrogen contents in both samples are within
the normal range for medical grade titanium alloys.